Detalle Publicación

ARTÍCULO

Application of the Monte Carlo efficiency transfer method to an HPGe detector with the purpose of environmental samples measurement

Autores: Morera Gomez, Yasser (Autor de correspondencia); Cartas-Aguila, H. A.; Alonso-Hernández, C. M.; Bernal-Castillo, J. L.; Guillén-Arruebarrena, A.
Título de la revista: APPLIED RADIATION AND ISOTOPES
ISSN: 0969-8043
Volumen: 97
Páginas: 59 - 62
Fecha de publicación: 2015
Resumen:
Monte Carlo efficiency transfer method was used to determine the full energy peak efficiency of a coaxial n-type HPGe detector. The efficiencies calibration curves for three Certificate Reference Materials were determined by efficiency transfer using a Eu-152 reference source. The efficiency values obtained after efficiency transfer were used to calculate the activity concentration of the radionuclides detected in the three materials, which were measured in a low-background gamma spectrometry system. Reported and calculated activity concentration show a good agreement with mean deviations of 5%, which is satisfactory for environmental samples measurement.